Simulation of low dose positron emission mammography scanner for global breast health applications

Journal Article


Download full-text (Open Access)

Abstract


  • Positron emission mammography ("PEM") is a breast imaging modality that typically involves the administration of relatively high doses of radiotracer. In order to reduce tracer costs and consider PEM for global screening applications, it would be helpful to reduce the required amount of administered radiotracer so that patient dose would be comparable to conventional x-ray mammograms. We performed GATE Monte Carlo investigations of several possible camera configurations. Increasing the detector thickness from 10 to 30 mm, increasing the camera surface area from 5×20cm2 to 20×20cm2, and applying depth-ofinteraction information to increase the acceptance angle, increased the overall efficiency to radiation emitted from a breast cancer by a factor of 24 as compared to existing commercial systems

UOW Authors


  •   Ryder, W. J (external author)
  •   Masoomi, M. A (external author)
  •   Wienberg, I. N (external author)
  •   Stepanov, P. S (external author)
  •   Urdaneta, M (external author)
  •   Anashkin, E (external author)
  •   Reznik, A (external author)
  •   Rosenfeld, Anatoly B.

Publication Date


  • 2010

Citation


  • Ryder, W., Masoomi, M., Wienberg, I., Stepanov, P., Urdaneta, M., Anashkin, E., Reznik, A. & Rozenfeld, A. (2010). Simulation of low dose positron emission mammography scanner for global breast health applications. N. Pelc & E. Samei In Medical Imaging 2010: Physics of Medical Imaging, 15-18 Feb, 2010, United States. Progress in Biomedical Optics and Imaging, 7622(49) 1-8.

Scopus Eid


  • 2-s2.0-84873920068

Ro Full-text Url


  • http://ro.uow.edu.au/cgi/viewcontent.cgi?article=6809&context=engpapers

Ro Metadata Url


  • http://ro.uow.edu.au/engpapers/3878

Number Of Pages


  • 7

Start Page


  • 1

End Page


  • 8

Volume


  • 7622(49)

Place Of Publication


  • Washington, USA

Abstract


  • Positron emission mammography ("PEM") is a breast imaging modality that typically involves the administration of relatively high doses of radiotracer. In order to reduce tracer costs and consider PEM for global screening applications, it would be helpful to reduce the required amount of administered radiotracer so that patient dose would be comparable to conventional x-ray mammograms. We performed GATE Monte Carlo investigations of several possible camera configurations. Increasing the detector thickness from 10 to 30 mm, increasing the camera surface area from 5×20cm2 to 20×20cm2, and applying depth-ofinteraction information to increase the acceptance angle, increased the overall efficiency to radiation emitted from a breast cancer by a factor of 24 as compared to existing commercial systems

UOW Authors


  •   Ryder, W. J (external author)
  •   Masoomi, M. A (external author)
  •   Wienberg, I. N (external author)
  •   Stepanov, P. S (external author)
  •   Urdaneta, M (external author)
  •   Anashkin, E (external author)
  •   Reznik, A (external author)
  •   Rosenfeld, Anatoly B.

Publication Date


  • 2010

Citation


  • Ryder, W., Masoomi, M., Wienberg, I., Stepanov, P., Urdaneta, M., Anashkin, E., Reznik, A. & Rozenfeld, A. (2010). Simulation of low dose positron emission mammography scanner for global breast health applications. N. Pelc & E. Samei In Medical Imaging 2010: Physics of Medical Imaging, 15-18 Feb, 2010, United States. Progress in Biomedical Optics and Imaging, 7622(49) 1-8.

Scopus Eid


  • 2-s2.0-84873920068

Ro Full-text Url


  • http://ro.uow.edu.au/cgi/viewcontent.cgi?article=6809&context=engpapers

Ro Metadata Url


  • http://ro.uow.edu.au/engpapers/3878

Number Of Pages


  • 7

Start Page


  • 1

End Page


  • 8

Volume


  • 7622(49)

Place Of Publication


  • Washington, USA