A simulation study of BrachyShade, a shadow-based internal source tracking system for HDR prostate brachytherapy

Journal Article


Abstract


  • This paper presents a simulation study of BrachyShade, a proposed internal source-tracking system for real time quality assurance in high dose rate prostate brachytherapy. BrachyShade consists of a set of spherical tungsten occluders located above a pixellated silicon photodetector. The source location is estimated by minimising the mean squared error between a parametric model of the shadow image and acquired images of the shadows projected on the detector plane. A novel algorithm is finally employed to correct the systemic error resulting from Compton scattering in the medium. The worst-case error obtained with BrachyShade for a 13.5 ms image acquisition is less than 1.3 mm in the most distant part of the treatment volume, while for 75% of source locations an error of less than 0.42 mm was achieved.

UOW Authors


  •   Alabd, Roumani (external author)
  •   Safavi-Naeini, Mitra (external author)
  •   Wilson, Keenan (external author)
  •   Rosenfeld, Anatoly B.
  •   Franklin, Daniel R. (external author)

Publication Date


  • 2018

Citation


  • Alabd, R., Safavi-Naeini, M., Wilson, K. J., Rosenfeld, A. B. & Franklin, D. R. (2018). A simulation study of BrachyShade, a shadow-based internal source tracking system for HDR prostate brachytherapy. Physics in Medicine and Biology, 63 (20), 205019-1-205019-15.

Scopus Eid


  • 2-s2.0-85055073400

Start Page


  • 205019-1

End Page


  • 205019-15

Volume


  • 63

Issue


  • 20

Place Of Publication


  • United Kingdom

Abstract


  • This paper presents a simulation study of BrachyShade, a proposed internal source-tracking system for real time quality assurance in high dose rate prostate brachytherapy. BrachyShade consists of a set of spherical tungsten occluders located above a pixellated silicon photodetector. The source location is estimated by minimising the mean squared error between a parametric model of the shadow image and acquired images of the shadows projected on the detector plane. A novel algorithm is finally employed to correct the systemic error resulting from Compton scattering in the medium. The worst-case error obtained with BrachyShade for a 13.5 ms image acquisition is less than 1.3 mm in the most distant part of the treatment volume, while for 75% of source locations an error of less than 0.42 mm was achieved.

UOW Authors


  •   Alabd, Roumani (external author)
  •   Safavi-Naeini, Mitra (external author)
  •   Wilson, Keenan (external author)
  •   Rosenfeld, Anatoly B.
  •   Franklin, Daniel R. (external author)

Publication Date


  • 2018

Citation


  • Alabd, R., Safavi-Naeini, M., Wilson, K. J., Rosenfeld, A. B. & Franklin, D. R. (2018). A simulation study of BrachyShade, a shadow-based internal source tracking system for HDR prostate brachytherapy. Physics in Medicine and Biology, 63 (20), 205019-1-205019-15.

Scopus Eid


  • 2-s2.0-85055073400

Start Page


  • 205019-1

End Page


  • 205019-15

Volume


  • 63

Issue


  • 20

Place Of Publication


  • United Kingdom